August 2021
2021-08-25
Authors: Massimo, Colin
- Working with the SiPM and the laser diode in the lab
- We don't have anything to digitize the signals with at the moment, so for now sending the SiPM signal direct to the oscilloscope
- Previously were having trouble reading the data out from the scope using a USB connection
- Hooked the scope up to the ethernet switch so we could get it on the network
- Currently: 10.44.45.11, but it will change if it gets switched off
- Connected with pyvisa -> open_resource('TCPIP::10.44.45.11::INSTR')
- Right now seeing some weird things in the externally triggered laser signals
- 1. a delay of around 20 ns between the two
- 2. oscillations in the signal (look like reflections)
2021-08-06
Author: Deivid R., Colin A.
- It does not work. Bad touchscreen. Rerouted VGA cable to a monitor to see how it’s behaving. SBC (the on-board computer) restarts once a minute on a loop, beeping once at the start. We checked bios and found nothing useful.
-
- Deivid was able to enter the BIOS using what was most likely the delete key (maybe the escape key was involved too)
- Seems like it uses Award BIOS / CMOS
https://thermotron.com/equipment/se-series-detail/se-600-3-3-environmental-chamber/





--
Deivid Ribeiro - 2021-08-06
Comments
*